Patient receives double dose of radiotherapy

By ThinkReliability Staff

The risk associated with medical treatment administration is high. There is a high probability for errors because of the complexity of the process involved in not only choosing a treatment, but ensuring that the amount and rate of treatment is appropriately calculated for the patient. The consequence associated with treatment errors is significant – death can and does result from inappropriately administered treatment.

Medical treatment includes delivery of both medication and radiation. Because of the high risk associated with administering both medication and radiation therapy, independent checks are frequently used to reduce risk.

Independent checks work in the following way: one trained healthcare worker performs the calculation associated with medical treatment delivery. If the treatment is then delivered to the patient, the probability that a patient will receive incorrect treatment is the error rate of that healthcare worker. (For example, a typical error rate for highly trained personnel is 1/1,000. If only one worker is involved with the process, there is a 0.1% chance the patient will receive incorrect treatment.) With an independent check, a second trained worker performs the same calculations, and the results are compared. If the results match, the medication is administered. If they don’t, a secondary process is implemented. The probability of a patient receiving incorrect treatment is then the product of both error rates. (If the second worker also has an error rate of 1/1,000, the probability that both workers will make an error on the same independently performed calculation is 1/1,000 x 1/1,000, or 0.0001%.)

However, in a case last year in Scotland, a patient received a significant radiotherapy overdose despite the use of independent checks, and verification by computer.   In order to better understand how the error occurred, we can visually diagram the cause-and-effect relationships in a Cause Map. The error in this case is an impact to the patient safety goal, as a radiotherapy overdose carries a significant possibility of serious harm. The Cause Map is built by starting at an impacted goal and asking “why” questions. All causes that result in an effect should be included on the Cause Map.

In this case, the radiotherapy overdose occurred because the patient was receiving palliative radiotherapy, the incorrect dose was entered into the treatment plan, and the incorrect dose was not caught by verification methods. Each of these causes is also an effect, and continuing to ask “Why” questions will develop more cause-and-effect relationships. The incorrect dose was entered into the treatment plan because it was calculated incorrectly (but the same) by two different radiographers working independently. Both radiographers made the same error in their manual calculations. This particular radiotherapy program involved two beams (whereas one beam is more common). The dose for each beam then must be divided by two (to ensure the overall dose is as ordered). This division was not performed, leading to a doubled calculated dose. The inquiry into the overdose found that both radiographers used an old procedure which was confusing and not recommended by the manufacturer of the software that controlled the radiotherapy delivery. While a new procedure had been implemented in February 2015, the radiographers had not been trained in the new procedure.

Once the two manual calculations are performed, the treatment plan (including the dose) was entered into the computer (by a third radiographer). If the treatment plan does not match the computer’s calculations, the computer sends an alert and registers an error. The treatment plan cannot be delivered to the patient until this error is cleared. The facility’s process at this point involves bringing in a treatment planner to attempt to match the computer and calculated doses. In this case, the treatment planner was one of the radiographers who had first (incorrectly) performed the dose calculation. The radiographers involved testified that alerts came up frequently, and that any click would remove them from the screen (so sometimes they were missed altogether).

The inquiry found that somehow the computer settings were changed to make the computer agree with the (incorrect) manual calculations, essentially performing an error override. The inquiry found that the radiographers involved in the case believed that the manually calculated dose was correct, likely because they didn’t understand how the computer calculated doses (not having had any training on its use) and held a general belief that the computer didn’t work well for calculating two beams.

As a result of this incident, the inquiry made several recommendations for the treatment plan process to avoid this type of error from recurring. Specifically, the inquiry recommended that the procedure and training for manual calculation be improved, independent verification be performed using a different method, procedures for use of the computer be improved (including required training on its use), and requiring manual calculations to be redone when not in agreement with the computer. All of these solutions will reduce the risk of the error occurring.

There is also a recommended solution that doesn’t reduce the risk of having an error, but increases the probability of it being caught quickly. This is to outfit patients receiving radiotherapy with a dosimeter so their received dose can be compared with the ordered dose. (In this case, the patient received 5 treatments; had a dosimeter been used and checked the error would likely have been noticed after only one.)

To view the Cause Map for this incident, please click on “Download PDF” above.

CDC provides guidance for states to respond to Zika cases

By ThinkReliability Staff

The first Zika cases related to the current outbreak were found in Brazil in May 2015, along with a dramatic increase in microcephaly in babies born in that year. (See our previous blog about the possible link – now verified – between Zika and microcephaly.) Microcephaly is a serious birth defect that impacts many children whose mothers contract Zika while pregnant.

Active Zika transmission currently exists in nearly all of South and Central America, the Caribbean, and some Pacific Islands. 934 people in the US have been infected with Zika; 287 of those infected are pregnant women. Most of these people were infected outside the country and then traveled to the US. Zika is primarily spread by mosquitos, but can also be transmitted through blood transfusion, laboratory exposure and sexual contact.

While no cases of transmission by mosquito have yet been reported in the continental US, the Centers for Disease Control and Prevention (CDC) has released a blueprint for states to respond to locally transmitted cases of Zika. A visual diagram outlining the steps to be taken from the blueprint (a Process Map) can be helpful. (To view the Process Map for the CDC’s interim Zika response process, click on “Download PDF”.)

The CDC’s plan involves four stages. The first stage is implemented during mosquito season. This stage involves surveillance for suspected locally transmitted infections (i.e. persons with “symptoms compatible with Zika virus infection who do not have risk factors for acquisition through travel or sexual contact”, with pending test results). Upon a suspected infection, state officials and the CDC should be notified. State or local officials will open an epidemiological investigation (including ongoing surveillance) and begin implementing controls, involving both reducing mosquito populations and continuing public outreach, with CDC assistance as needed.

Stage 2 occurs upon confirmation of a locally transmitted infection. At this point, notification expands to include local blood centers as well as others required by International Health Regulations. The CDC will assist with an expanded investigation, surveillance, and communication, including deployment of an emergency response team (CERT) if desired. Once Stage 2 has been reached, stand down will only occur after 45 days (3 mosquito incubation periods) without additional infections or when environmental conditions no longer permit transmission.

If there is confirmed Zika in two or more persons whose movement during the exposure period overlaps within a one-mile diameter, Stage 3 (widespread local transmission) is entered. First, local officials will attempt to determine the transmission area, the “geographic area in which multiperson local transmission has occurred and may be ongoing”. Communication, surveillance, testing and controls are enhanced and expanded. Interventions for blood safety and vulnerable populations (including pregnant women) are implemented.

Once the infection has spread outside a county, it enters Stage 4 (widespread multijurisdictional transmission). All steps taken in previous stages are expanded and enhanced. The CDC will evaluate whether local capacity is adequate for response, and will assist as needed. Stage 4 actions will be continued until the criteria for stand down is met.

Based on previous experience with two mosquito-transmitted diseases, chikungunya & dengue fever, the CDC does not believe Stage 4 will be reached within the United States. However, as Dr. Tim F. Jones, an epidemiologist for the State of Tennessee, says, “Even though the percentages and the likelihoods are incredibly low, the outcome is awful.” Risk is a function of probability and consequence. Even with a low probability, the high consequence makes the risk from Zika considerable, and worth planning for.

To view the Process Map, click on “Download PDF” above. Or, click here to view the CDC’s interim guidance.

.